Direct Determination of Arsenic in Potassium Citrate Tablet Using Graphite Furnace Atomic Absorption Spectrometry

AuthorMohsen Shahlaeien
AuthorAlireza Pourhosseinen
Issued Date2012-06-30en
AbstractIn this study, a rapid and simple method based on graphite furnace atomic absorption spectrometry (GFAAS) is described for direct determination of arsenic in potassium citrate tablet samples. Pyrolytic graphite tubes were used as atomizers. Nickel nitrate was used as chemical modifier. The ashing and atomization temperature and the concentration of modifier were also optimized. The detection limit of the method was 0/8 ng/ml As in potassium citrate tablet samples.The relative standard deviation for ten determination of a spiked sample with concentration of 30 ng/ml As ranged was 2.2 %. The accuracy of the method was confirmed by the analysis of spiked samples. The linear rang of calibration is in the range of 0.8-80 ng/ml of arsenic. en
DOIhttps://doi.org/en
URIhttps://brieflands.com/journals/jrps/articles/147798en
KeywordArsenicen
KeywordPotassium Citrate Tableten
KeywordETAASen
PublisherBrieflandsen
TitleDirect Determination of Arsenic in Potassium Citrate Tablet Using Graphite Furnace Atomic Absorption Spectrometryen
TypeResearch Articleen

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
direct_determination_of_arsenic_in_potassium.3.pdf
Size:
221.13 KB
Format:
Adobe Portable Document Format
Description:
Article/s PDF