Development of a Simple RP-HPLC-UV Method for Determination of Azithromycin in Bulk and Pharmaceutical Dosage forms as an Alternative to the USP Method

AuthorTayebeh Gharien
AuthorFarzad Kobarfarden
AuthorSeyed Alireza Mortazavien
Issued Date2013-03-31en
AbstractThe present study was designed to develop a simple, validated liquid chromatographic method for the analysis of azithromycin in bulk and pharmaceutical dosage forms using ultraviolet detector. The best stationary phase was determined as C18 column, 5 μm, 250 mm × 4.6 mm. Mobile phase was optimized to obtain a fast and selective separation of the drug. Flow rate was 1.5 mL/min, Wavelength was set at 210 nm and the volume of each injection was 500 μL. An isocratic methanol/buffer mobile phase at the ratio of 90:10 v/v gave the best separation and resolution. The proposed method was accurate, precise, sensitive, and linear over a wide range of concentration of azithromycin. The developed method has the advantage of using UV detector compared to the USP method in which electrochemical detector has been used. The validated method was successfully applied to the determination of azithromycin in bulk and pharmaceutical dosage forms.en
DOIhttps://doi.org/10.22037/ijpr.2013.1272en
KeywordAzithromycinen
KeywordHPLCen
KeywordUV detectionen
KeywordDeterminationen
KeywordPharmaceuticalsen
PublisherBrieflandsen
TitleDevelopment of a Simple RP-HPLC-UV Method for Determination of Azithromycin in Bulk and Pharmaceutical Dosage forms as an Alternative to the USP Methoden
TypeOriginal Articleen

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ijpr-12-57.pdf
Size:
745.17 KB
Format:
Adobe Portable Document Format
Description:
Article/s PDF