Determining the Correlation and Accuracy of Three Methods of Measuring Neonatal Bilirubin Concentration

AuthorMirhadi Mussavien
AuthorPedram Niknafsen
AuthorBahareh Bijarien
Accessioned Date2023-12-24T16:01:49Z
Available Date2023-12-24T16:01:49Z
Issued Date2013-06-30en
AbstractObjective: There are different methods for measuring bilirubin concentration; however, it is quite important for practitioners to know which method should be used in certain clinical situations. The present prospective study aimed to compare three different methods for measuring neonatal bilirubin concentrations. Methods: All full term neonates who were either brought into emergency departments or admitted to the neonatal wards in Kerman city in 2011 were recruited (n=428). The correlation coefficients were estimated for the routine ways of bilirubin concentrations including "Capillary", "Cutaneous" and "Laboratory" methods. Findings: Of 428 recruited neonates, 178 were female. Mean age ±SD was 178±71 hours. The correlation coefficient for "David Icterometer" vs "JM103" was 0.91, while the corresponding coefficient for "David Icterometer" vs "Capillary" was 0.96. It was also equivalent to 0.85 for correlation between "JM103" and "Capillary" methods. The David Icterometer measured an average of 2.36 mg/dl levels of bilirubin concentration compared to the JM103 method. The Capillary method showed a lower bilirubin level than the venous concentration (0.91 mg/dl on average). Compared with the "Capillary", the "JM103" measured a slightly higher level of bilirubin with an average 0.57 mg/dl. Conclusion: Due to low difference (less than 1 mg/dl) between "JM103" and the "Capillary methods" for measurement of neonatal bilirubin concentration, these two methods could alternatively be used instead of usual laboratory method.en
DOIhttps://doi.org/en
URIhttps://repository.brieflands.com/handle/123456789/35080
KeywordCapillary Bilirubinometeren
KeywordNeonateen
KeywordBilirubinen
KeywordDavid Icterometeren
PublisherBrieflandsen
TitleDetermining the Correlation and Accuracy of Three Methods of Measuring Neonatal Bilirubin Concentrationen
TypeResearch Articleen

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
IJPD-23-333.pdf
Size:
346.2 KB
Format:
Adobe Portable Document Format
Description:
Article/s PDF