Decay-Delay Technology for Management of Defibrillator T-wave Oversensing

AuthorAmir Aslanien
AuthorM Moradien
AuthorJ Kheirkhahen
AuthorM Kafien
Issued Date2010-12-31en
AbstractAppropriate sensing is indispensable for appropriate functioning of implantable cardioverter-defibrillator (ICD). Oversensing is a major clinical dilemma in patients with ICD leading to inappropriate ventricular arrhythmia detection and therapy. T-wave oversensing is an important clinical problem leading to QRS double counting with inappropriate therapy, in particular. We present a patient with dilated cardiomyopathy and intermittent T wave oversensing by an ICD which could be resolved by device reprogramming.en
DOIhttps://doi.org/en
Keywordaen
PublisherShiraz University of Medical Sciencesen
TitleDecay-Delay Technology for Management of Defibrillator T-wave Oversensingen
TypeCase Reporten