Decay-Delay Technology for Management of Defibrillator T-wave Oversensing
Author | Amir Aslani | en |
Author | M Moradi | en |
Author | J Kheirkhah | en |
Author | M Kafi | en |
Issued Date | 2010-12-31 | en |
Abstract | Appropriate sensing is indispensable for appropriate functioning of implantable cardioverter-defibrillator (ICD). Oversensing is a major clinical dilemma in patients with ICD leading to inappropriate ventricular arrhythmia detection and therapy. T-wave oversensing is an important clinical problem leading to QRS double counting with inappropriate therapy, in particular. We present a patient with dilated cardiomyopathy and intermittent T wave oversensing by an ICD which could be resolved by device reprogramming. | en |
DOI | https://doi.org/ | en |
Keyword | a | en |
Publisher | Shiraz University of Medical Sciences | en |
Title | Decay-Delay Technology for Management of Defibrillator T-wave Oversensing | en |
Type | Case Report | en |
Files