A Novel High Performance Liquid Chromatographic Method for Determination of Nystatin in Pharmaceutical Formulations by Box–Behnken Statistical Experiment Design

AuthorFarnaz Shokranehen
AuthorRamin Asgharianen
AuthorAssem Abdollahpouren
AuthorMehdi Raminen
AuthorAli Montaserien
AuthorArash Mahboubien
Issued Date2015-03-31en
AbstractIn this study a novel High Performance Liquid Chromatography for the assay of nystatin in oral and vaginal tablets were optimized and validated using Box–Behnken experimental design. The method was performed in the isocratic mode on a RP-18 column (30 °C) using a mobile phase consisting of ammonium acetate 0.05 M buffer/ Methanol mixture (30:70) and a flow-rate of 1.0 mL/min. The specificity, linearity, precision, accuracy, LOD and LOQ of the method were validated. The method was linear over the range of 5–500 µg/mL with an acceptable correlation coefficient (r2 = 0.9996). The method’s limit of detection (LOD) and quantification (LOQ) were 0.01 and 0.025 µg/mL respectively. The results indicate that this validated method can be used as an alternative method for assay of nystatin.en
DOIhttps://doi.org/10.22037/ijpr.2015.1711en
KeywordNystatinen
KeywordHigh performance liquid chromatographyen
KeywordBox–Behnken experimental designen
PublisherBrieflandsen
TitleA Novel High Performance Liquid Chromatographic Method for Determination of Nystatin in Pharmaceutical Formulations by Box–Behnken Statistical Experiment Designen
TypeOriginal Articleen

Files