Myocardial infarction caused by active-fixation atrial pacing lead screwed into the anterior descending coronary artery

AuthorEglė Majauskienėen
AuthorGediminas Aucinaen
AuthorPalmyra Semenieneen
AuthorPaulius Jurkuvenasen
AuthorPranas Serpytisen
OrcidEglė Majauskienė []en
OrcidGediminas Aucina []en
OrcidPalmyra Semeniene []en
OrcidPaulius Jurkuvenas []en
OrcidPranas Serpytis []en
Issued Date2018-12-31en
AbstractCoronary artery injury is a particularly rare complication of pacemaker implantation. Herein, we reported the case of a 76-year-old woman with sick sinus syndrome undergoing permanent pacemaker implantation who presented with chest pain a day after the procedure. During angiography, a pacemaker lead screwed into the left anterior descending coronary artery was found. The coronary artery was covered by a stent and the lead position was corrected. Based on the published research, the majority of right ventricular pacemaker leads implanted on the septum are indeed implanted on the anteroseptal junction. Since, the left anterior descending coronary artery covers this site, the risk of artery injury is higher.en
DOIhttps://doi.org/en
KeywordPacemakeren
KeywordMyocardial Infarctionen
PublisherBrieflandsen
TitleMyocardial infarction caused by active-fixation atrial pacing lead screwed into the anterior descending coronary arteryen
TypeCase Reporten